2. Insert a Flash Integrator through the open exhaust port and position it under the basket or rack near the exhaust port, allowing the end of the integrator to extend out through the opening.
3. Position the Flash-Guard® in the autoclave with the exhaust port in the fully open position, facing out toward the operator. Slide the lid forward so at least a three (3) inch opening is created on the rear portion of the Flash-Guard® unit. This is essential for proper steam circulation. Proceed with flash sterilization utilizing recommended guidelines and following autoclave manufacturer's operator instructions for flash sterilization.
4. When the sterilization cycle is complete, open the sterilizer, remove and read the integrator and place the integrator onto the lid. SLIDE LID CLOSED & CLOSE EXHAUST DOOR: Using the universal handle, slide the lid to the closed position. Then, swing the exhaust door closed - the door will be ready to be snapped shut by using the rubber end of the handle to latch the door completely.
5. Then, remove the unit from the autoclave, sliding it onto the transfer cart. Lifting the unit is not recommended or necessary; The universal handle can also be used to pull the container onto a cart. The Flash-Guard® unit is now ready for sterile transport to the point of use.